SpeXor BIG

Technical specifications

Unit of measure millimeters / inches (on request)
Measurement system Hall effect sensor
Display Liquid crystal digital LCD
Power supply 5V USB cable
Usage temperature +5 °C ~ +40 °C
Range 0-20 mm

Preliminary operations

  1. Clean the surfaces to be measured.
  2. Clean the feeler gauge, in particular the sphere and the measuring tip.
  3. Bring the sphere closer to the tip so that it touches the flat part with the sensor (zero value).
  4. Hold the tip with the ball in a vertical position.
  5. Switch on the thickness gauge by connecting the power cable.
  6. Wait for the end of the initial calibration.
  7. Carry out some measurements with no load. A flashing W means Wait; after a few flashes an @ appears, indicating that it is calibrated.
  8. For a more accurate measurement, work at room temperature, carry out some measurements with no load on a thickness of ~ 10 mm and wait at least 5 minutes.

Measurements

  1. Place the sphere on the surface to be measured; on the opposite surface, in correspondence with the sphere, place the tip.
  2. By magnetism the sphere is attracted to the tip without the need for forcing.
  3. Read the measured value.
  4. Extract the sphere from the piece under examination.
  5. Turn off SpeXor by unplugging the power cord.
  6. After taking the measurement and without moving the tip, it is possible to reset (0.00) the measured value by pressing the button. The subsequent measurement will detect the difference between the first and second measurement (positive or negative with sign - before the value).

The magnetic attraction force between the tip and the sphere may not be sufficient to prevent the latter from falling. Pay attention.

SpeXor is very precise in the range from 0 to 5 mm, where it is possible to stay in the tenth of a mm with appropriate precautions. Make sure it self-calibrates before each measurement: an @ appears when recalibrating. Work at room temperature if possible.

Warranty: 24 months

How to measure